4.4.4.2 The audit team conducts the review based on applicable standards.
4.4.5 Generate audit report.
4.4.5.1 After each audit, the audit team leader shall prepare the "Audit Report" based on relevant information identified during the on-site audit.Report," and is responsible for its content. Upon approval by the Technical Committee, it shall be issued to the certification applicant.
4.4.5.2 The report shall provide an accurate, concise, and clear record of the audit to support certification decisions. It shall include the following:
a) Customer name, address, and management representative;
b) Audit Type (e.g., initial certification, surveillance, or recertification audit)
c) Purpose, scope, and criteria of the audit;
d) Audit team members and audit schedule;
e) Statement of compliance with applicable certification requirements;
f) Time period covered by the report;
g) Non-conformance situations;
h) Audit conclusion.
4.5Corrective actions for nonconformities and verification of their results
4.5.1For non-conformities identified during the audit, the organization shall require the applicant to analyze the root cause and take corrective actions within Take corrective action within the specified timeframe.
4.5.2 The organization shall verify the effectiveness of corrective actions and measures taken by the auditee.
4.6Authentication Decision
4.6.1This organization should comprehensively evaluate the audit report, nonconformities, corrective actions and their results. Based on this, make the certification decision.
The audit team members for 4.6.2 shall not participate in the certification decision for the audit project.
4.6.3 Before making a certification decision, the organization shall confirm the following:
a. The audit report complies with Rule 4.4. The requirements necessary to provide the information needed to make a certification decision.
b. The institution has reviewed, accepted, and verified the effectiveness of corrective actions and their results for nonconformities reflecting the following issues.
Failed to meet requirements of the medical device quality management system standard.
2. Management objectives are not measurable, or the measurement methods are unclear.
③ Monitoring and measurement of key points that have a significant impact on achieving management objectives are not functioning effectively, or these key pointsIncomplete or invalid report or review records.
④ Deficiencies exist in the effectiveness of continuous improvement of the medical device quality management system, raising significant concerns about achieving management objectives.
c. The organization has reviewed other nonconformities and accepted the applicant's plan for corrective actions.
4.6.4 meets 4.6.3 If the organization provides sufficient objective evidence that it meets the following requirements, the certification body shall determine that the organization complies with the certification requirements and issue a certificate.
a. The applicant's medical device quality management system complies with the required standards and operates effectively.
b. Products or services covered by the certification comply with applicable laws and regulations.
c. The applying organization has fulfilled its obligations as specified in the certification contract.
4.6.5If an organization's application does not meet the above requirements, it shall be deemed non-compliant with certification requirements and notified in writing. Submit an application and explain the reasons for failing certification.
4.6.6After issuing a certification certificate, this organization shall 30 within working days, report the relevant information in accordance with prescribed requirementsSend to CNCA. CNCA on its website(www.cnca.gov.cn)and the company website (http://www.zjsjjc.com)Publish the column publicly to societyCertification certificate information submitted by the organization.
4.6.7This organization shall not link the certification status of the applicant organization to the compensation of auditors and other personnel involved in the certification audit.linked to compensation.
5Surveillance Audit Procedure
5.1 The organization shall effectively monitor and supervise certified organizations (hereinafter referred to as "certificate holders") that hold its medical device quality management system certification certificates, ensuring that their certified quality management systems remain compliant.
5.2 To ensure 5.1 is achieved The requirements are that the organization shall, based on the medical device quality risk level of the certified organization or other...Features: Determine the frequency of surveillance audits for certified organizations.
5.2.1 Surveillance audits shall be conducted at least once per calendar year, except for years in which recertification is required. The first surveillance audit following initial certification shall take place within [X] months from the date of the certification decision. 12 within the month.
Note: The frequency of surveillance audits may need to be adjusted to account for factors such as seasonality or time-limited management system certifications (e.g., temporary construction sites).
5.2.2 When the deadline for the second surveillance audit is reached, but there is evidence that the certified organization is temporarily unable to undergo the surveillance audit.When applicable, the interval for surveillance audits may be extended, but shall not exceed 15 months.
If surveillance audits under 5.2.3 are not conducted within the specified period, they shall be handled in accordance with clauses 7.2 or 7.3.
The 5.3 surveillance audit timeline is calculated per company documentation.
The audit team for the 5.4 surveillance audit shall meet the requirements of clause 4.3.2.
5.5 Surveillance audits may be conducted on-site or remotely at the certified organization. Due to seasonal production cycles, it may not be feasible to cover all products during each surveillance audit; however, throughout the certificate's validity period, surveillance audits must cover the full scope of certification.all products within.
At least the following items must be reviewed during 5.6 supervision and audit:
a. Have there been any changes to the activities covered by the medical device quality management system and the resources of the operating system since the last audit?
b. Whether key critical points are operating normally and effectively in accordance with the requirements of the medical device quality management system.
c. Are the corrective and preventive actions taken for non-conformities identified in the previous audit still effective?
d.The activities covered by the medical device quality management system that are subject to legal and regulatory requirements must remain in continuous compliance with such regulations.
e. Achievement of objectives. Where applicable, if objectives were not met, the certified organization must investigate and analyze the causes and implement corrective actions during internal management reviews in a timely manner.
f. Whether the certified organization's use of the certification mark or reference to its certification status complies with relevant regulations.
g. Whether internal audits and management reviews are conducted in a standardized and effective manner, as applicable.
h. Timely acceptance and handling of complaints.
i.When applicable, timely effective continuous improvements were implemented in response to issues identified during internal audits or reported through complaints.
5.7The audit report for surveillance audits shall be5.6 Describe audit evidence and audit findings for each specified audit requirement.and audit conclusion. The audit team shall provide recommendations on whether to maintain the certification certificate.
5.8 The institution will decide to maintain, suspend, or revoke the certification certificate based on the surveillance audit report and other relevant information.
6Reauthentication Process
Before the 6.1 certification expires, if the certified organization applies to retain its certificate, the certification body shall conduct a recertification audit and decide whether to renew the certificate.
6.2 Certification Authority shall4.3.2A review team is formed by {0} requirements. Following4.3Develop a recertification audit plan based on the requirements and past surveillance audit results, and submit it to the audit team for implementation. The audit team shall carry out the recertification audit in accordance with the plan.
When there are no significant changes to the internal and external environment of a certified organization's medical device quality management system, the Stage 1 audit for recertification may be omitted, provided that the total audit time is not less than4.3.1Calculating person-days: 2/3.
For serious nonconformities identified during the recertification audit, the certification body shall set a deadline for the certified organization to implement corrective actions and corrections, and complete verification of these actions before the original certificate expires.
The certification body 6.4 makes the recertification decision based on 4.6 requirements. If the certified organization continues to meet certification requirements and fulfill its contractual obligations, a new certificate shall be issued.
6.5If recertification activities are completed before the current certification certificate's expiration date and you choose to issue a new certificate, the new certificate's expiration date may be based on the current certificate's expiration date. The issuance date on the new certificate must not be earlier than the recertification decision date.
If the certification body fails to complete the recertification audit or address serious nonconformities before the current certificate's expiration date,If corrective actions and corrections are not verified, re-certification shall not be granted, and the validity of the original certification certificate shall not be extended. If, after the current certification certificate expires, the certification body is able to6Complete any outstanding recertification activities within {0} months to restore certification; otherwise, a second-stage audit must be conducted at minimum. The certification certificate's effective date shall not be earlier than the recertification decision date, and its expiration date shall be based on the previous certification cycle.
7Suspend or revoke certification
7.1 The organization shall establish and document procedures for suspending, revoking certification, or reducing the scope of certification.
7.2 Suspend Certificate
The certification body shall suspend the certificate of 7.2.1 within 5 working days after investigation and verification if any of the following circumstances occur.
a. The medical device quality management system fails to meet certification requirements on an ongoing or serious basis, including requirements for the effective operation of the system.
b. Fails to assume or fulfill the responsibilities and obligations stipulated in the certification agreement.
c. Ordered to suspend operations for rectification by relevant law enforcement or regulatory authorities.
d. Certificate suspension required due to identified issues in system operation by local certification regulatory authorities.
e. The administrative licensing proofs, qualification certificates, mandatory certification certificates, etc., held have expired or become invalid; the resubmitted application has been accepted but the certificate renewal has not yet been completed.
f. Voluntarily requested suspension.
g. Other circumstances requiring suspension of the certification certificate.
The suspension period for 7.2.2 certification shall not exceed 6.months. But belonging to7.2.1th e The suspension period for such cases may extend until the relevant authority issues a decision on the permit.If the certification authority is able to6complete remaining tasks within the monthMonitor and AuditThe activity, then the certification can be restored.
7.2.3Information regarding the suspension of certification certificates by this organization shall clearly state the suspension start date and duration, and declare that during the suspension... Certified organizations shall not use the certification certificate, certification mark, or reference certification information during this period.
7.3 Revoke Certificate
If a certified organization under 7.3.1 falls under any of the following circumstances, this body shall take action within 5 working days after obtaining and verifying the relevant information. Revoke its certification certificate today.
a. Documents proving the cancellation or revocation of legal status.
b. Refuses to cooperate with supervisory and inspection activities conducted by the certification regulatory authority, or provides false materials or information in response to inquiries or investigations regarding relevant matters.
c.A major medical device quality liability incident occurred, confirmed by the law enforcement regulatory authority to be caused by violations committed by the certified organization.
d. Has committed other serious violations of laws or regulations.
e. The suspension period for the certification certificate has expired, but the issue causing the suspension has not been resolved or corrected (including held administrative licenses)Proofs, qualification certificates, mandatory certification certificates, etc., have expired or become invalid, but the application was not approved.
f. The medical device quality management system is not in operation or no longer meets operational requirements.
g. Failure to properly cite or promote obtained certification information in accordance with relevant regulations, resulting in serious adverse effects or consequences, or failure to correct the issue within 6 months after being required to do so by this institution.
h. Other circumstances warranting the revocation of the certification certificate.
7.3.2After revoking a certification certificate, the organization shall promptly retrieve the revoked certificate. If retrieval is not possible, the organization shallThe decision to revoke shall be published or announced on relevant media and websites.
7.4The institution shall publish relevant information on its website when suspending or revoking certification certificates, and follow the prescribed procedures and requirements. Submit to the National Certification and Accreditation Administration.
7.5 This institution is obligated and responsible for taking effective measures to prevent the continued use of invalid certification certificates and marks. Use.
8Certification Requirements
The 8.1 certification must include at least the following information:
a. Certified organization name, address, and organizational code. This information must match the details in the legal status documentation.
b. Scope of business activities for production, operation, or services covered by the medical device quality management system. If the certified medical device quality management system covers multiple locations, provide the names and addresses of the relevant locations; this information must match the details in the corresponding legal status documents.
c. Medical Device Quality Management SystemGB/T 42061-2022/ISO 13485:2016 Compliant.
d. Certificate Number.
e. Institution name.
f. Certificate issue date and validity period (start and end dates). For recertification certificates issued without interruption since initial certification, the date of initial certification may be stated.
g. Certificate Verification Methods. In addition to publishing the certificate verification method on our website, we must also include it on the certificate itself. Note Clear :“ This Certificate Book Message Info Yes In Country Home Recognize Certificate Recognize Yes Monitor Supervise manage Process Commit Employee Yes Official Fang Net site
(www.cnca.gov.cnquery on", for public oversight.
8.2 The certification is valid for 3 years.
8.3 The organization shall establish a certificate information disclosure system. In addition to providing certification certificate information to applicant organizations and regulatory authorities, the organization shall also provide such information upon request from relevant social stakeholders to ensure public oversight.
9Integrated audit with other services and management systems
When conducting combined audits of medical device quality management systems and other management systems, 9.1, general or common requirements must comply with these rules. The audit report shall clearly reflect the requirements of clause 4.4 and ensure they are easily identifiable.
When combining audits with 9.2, the number of audit days for the combined audit should be determined based on the company's audit schedule.
10Process Conversion Certification
10.1 This institution must fulfill its social responsibilities and strictly prohibit accepting non-compliant cases for profit.GB/T 42061-2022/ISO 13485:2016 Organizations that do not effectively implement the medical device quality management system in accordance with standards are ineligible to convert their certification.
10.2 This organization shall accept applications to convert its certification certificates, thoroughly understand the reasons for conversion, and conduct necessary on-site audits.
10.3 Conversion applies only to currently valid certification certificates. Applications for conversion will not be accepted for certificates that are suspended, under suspension or revocation proceedings, or have expired.
10.4 Ordered to suspend operations or listed on the "blacklist" by regulatory authorities (e.g. 7.2 Article c item), issuedThe institution has revoked the certificate (e.g.,7.3 unless the organization undergoes comprehensive rectification, resulting in suspension or revocation of certification.The certificate issue has been resolved; otherwise, the certification application should not be accepted.
11 Accept organization appeals
When a certified organization disagrees with the certification decision, our institution shall accept its appeal and process it promptly.The written notification of the handling results shall be delivered to the certified organization within 60 days.
The written notice shall inform the certified organization that, if it believes the certification body has failed to comply with applicable certification laws, regulations, or rules.and cause self-combinationIf your legal rights are seriously infringed, you may directly file a complaint with the local certification regulatory authority or the Certification and Accreditation Administration of China.
Manage 12 authentication records
12.1 The institution shall establish a certification record retention system to document and securely preserve all phases of the certification process.
12.2 Records must be accurate and authentic to demonstrate that certification activities are effectively implemented. Records shall be in Chinese and retained for at least the duration of the certificate's validity period.
When records are saved electronically as 12.3, an non-editable electronic document format shall be used.
13Other
13.1 Mentioned in the rulesGB/T 42061-2022/ISO 13485:2016Standard time refers to the time when the certification activity occurs.This is the effective version of the standard. When describing the standard number in certification activities and certificates, use the complete standard number of the version that was effective at that time.
13.2 Copies of all documents referenced in this rule must be reproduced from the original and signed/stamped by the provider to confirm they match the original.
13.3 This organization can take necessary measures to conduct training and dissemination on the medical device quality management system and relevant technical standards, ensuring all employees correctly understand and implement these standards.